Precise QRS detection is an important first step in the analysis of heart rate variability. Algorithms based on the differential ECG are computationally efficient and therefore ideal for the real-time analysis of large datasets. Here, we discuss traditional first-derivative-based quadrature function (Hamilton-Tompkins) and Hilbert transformation-based methods of QRS detection and their modifications with improved detection thresholds. In a standard ECG dataset, the Hamilton-Tompkins algorithm had the highest detection accuracy (99.68% sensitivity, 99.63% positive prediction), but also the greatest time error. The modified Hamilton-Tompkins algorithm, as well as algorithms based on the Hilbert transformation, had comparable but slightly lower accuracy; However, these automated algorithms present an advantage for real-time applications by avoiding human intervention in threshold determination. The high accuracy of the method based on the Hilbert transform as compared to detection with the second ECG derivative is attributable to its intrinsically uniform spectrum of magnitude. For all algorithms, detection errors occurred mainly in beats with decreased signal slope, such as wide arrhythmic beats or attenuated beats. To obtain the best performance, a combination of the quadrature function and the algorithms based on the Hilbert transformation can be applied such that differences in detection point to signal abnormalities that can be further analyzed.