Technique for measuring the Dielectric constant using open ended coaxial cable in the
#1



[attachment=8010]

Ajaya Kumar Kavala
Research Scholar
Department of Physics
IIT Patna


Introduction of Dielectrics

Dielectric is an entity ,in which equal positive and negative charges separated by equal small distance.
In which each molecule has permanent dipole moment P
This type of materials are known as polar materials
Different molecules has different different directions of the dipole moment.
In Non polar materials ,The centre of the positive charge distributution in an atom or molecules coincides with the centre of the negative charge distribution.
The atoms or molecules do not have permanant dipole moment
Dielectric constant = capacitence without dielectric over capacitence with dielectric

Types of polarizations

Definition: Dipole moment per unit volume is called Polarisation
3 types of polarisation
1.Dipole polarisation
2.Ionic polarisation
3.electronic polarisation

Principle

An electromagnetic wave of particular frequency passed through a coaxial probe to the sample, some part of the wave is transmitted and some part of the wave reflected.

From the reflected signal we can get the reflection coefficient of that sample, from that we can get the corresponding impedance of that sample at that particular frequency

From that impedance we can obtain the complex dielectric constant at that frequency
Accurate determination of the frequency-dependent probe-end impedance in vacuum .
Elimination of spurious impedances such as those due to connector mismatch
Indeterminacy of probe parameters
Accurate modeling of the probe-liquid interface impedance after the above problems are eliminated.


Experimental data collection is described below step by step
Open
With the coax terminated by free space, the measurement plane of the ANA was moved to the coax end using the electrical delay provided. The delay, which corresponds to the length of the coax (typically 6-12 in), was adjusted to give a cluster of points near the real and imaginary parts.

A short at the coax end was created by raising a small vessel (about 6 cc) filled with mercury, until the coax end was well within the liquid. This resulted in a cluster of points around the position [Re and Img parts.
Standard Liquid
At this point, the display was checked to ensure that the date returned to the configuration for an open. A cell (of volume typically <10 cc) with a standard liquid, usually acetone, was now inserted so that the coax end was well immersed in the liquid. The data ρc were again read into the computer.




Reply

Important Note..!

If you are not satisfied with above reply ,..Please

ASK HERE

So that we will collect data for you and will made reply to the request....OR try below "QUICK REPLY" box to add a reply to this page
Popular Searches: ppt dielectric relaxation, dielectric constant experiment viva question, high k dielectric ppt, peer to peer network coaxial cable, coaxial cable topic for seminar, dielectric constant viva question with answers in pdfstract, integration constant,

[-]
Quick Reply
Message
Type your reply to this message here.

Image Verification
Please enter the text contained within the image into the text box below it. This process is used to prevent automated spam bots.